Abstract
To assess feasibility, image quality and measured venous caliber of non-contrast MRV (NC-MRV) of central and upper extremity veins, compared to contrast-enhanced MRV (CE-MRV) and ultrasound (US) in healthy volunteers.
10 subjects underwent NC-MRV and CE-MRV at 1.5 T, with comparison to US. Two radiologists evaluated MRI for image quality (IQ) and venous caliber.
NC-MRV is feasible, with inferior IQ but comparable venous caliber measurements CE-MRV (mean 7.9±4.58 mm vs. 7.83±4.62, p=0.13). Slightly larger upper limb caliber measurements were derived for NC-MRV and CE-MRV compared to US (NC-MRV 5.2±1.8 mm, CE-MRV 4.9±1.6 mm, US 4.5±1.8 mm, both p<0.001).